Jensprima online hardness and residual chlorine analyzers play a vital role in the service life of reverse osmosis membranes and the water recovery rate within pharmaceutical RO systems.
LEARN DETAILS
Online Water Hardness Analyzer PACON 5000
Online Water Hardness Analyzer PACON 4800
Online Water Hardness Analyzer PACON 4200
Total Hardness Reagent TH5000-TH5100
Alkalinity Analyzer PACON 4800
SCD Analyzer Flumsys 10TC-SP
Residual / Total Chlorine Analyzer PACON 2501
Residual Chlorine Analyzer FlumPANEL 2501
pH/Redox Controller InnoCon 6501P
pH Electrode InnoSens 125T
Conductivity InnoSens 350-0.01
Suspended Solids Sensor Innosens 815S
Ammonia Nitrogen Controller 10TC-I3
Nitrate Nitrogen Electrode InnoSens 560
Electromagnetic Flowmeter InnoMag 300




Table of contents
1. Failed to reset the remaining reagent level after installing or replacing new reagents.
2. Alarm persists after reset and restart; replace the SD card directly.

How should the analyzer handle an alarm triggered by excessively low reagent levels (<10%)?
The solution is as follows:
3.First, with the instrument in a non-measuring state, press and hold the [?] button for more than 3 seconds. A prompt as shown in the figure will appear. Select “OK”.

4.When the automatic chemical dosing countdown begins, observe whether any chemical is flowing out of the treatment tank. Wait until there is no air in the chemical line before pressing the “OK” button.

5、Select “Yes” and press the “OK” button to confirm.

6、After returning to the main interface, the remaining reagent level will reset to 100%.