multi parameter water quality analyzer
lron Ions Analyzer
PACON 3400
HOME PRODUCS PARAMETERS Iron lron Ions Analyzer lron Ions Analyzer PACON 3400


lron Ions Analyzer PACON 3400
lron Ions Analyzer PACON 3400

lron Ions Analyzer PACON 3400


PACON 3400 is a compact, easy-to-operate and accurate on-line iron ion analyzer from JENSPRIMA that measures the concentration of dissolved iron (divalent and trivalent iron) in water for controlled removal of iron ions. Using the colorimetric principle, it is very easy to switch between the LRS (0.01-0.5 mg/L) and HRS (0.2-6.0 mg/L) measurement ranges at any time by using different coloured LED light sources and selecting reagents.



Product Features:


1、Durable, accurate analyzer method

2、Automatic on-line monitoring, only one reagent required

3、Special material measuring tank, condensation will not adhere to the optics

4、Automatic zero point adjustment before each measurement to ensure measurement stability

  • Measurement MethodColourimetric method
    Measuring RangeLow range:0.01-0.5mg/L、High range:0.2-6.0mg/L
    Accuracy± 10%
    Repeatability± 5%
    Reagent Consumptionca. 0.5 ml/analysis 
    Expiry Date of Reagent1 year (<25°c, storage="" in="" shade="">
    Water Consumptionapprox. 2 L of water per analysis (at 2 bar pressure)
    Power Supply85 – 265 VAC, 47-63Hz
    Power Consumption25VA (operating), 3.5VA (standby)
    Protection ClassIP65
    DisplayGraphic backlit LCD display
    Measurement Unitsmg/L
    Outputs1、4 sets of programmable relay outputs (max. 250 V, 4 A)
    2、1 group of 0 / 4 – 20 mA signals, max. 750 Ω
    3、RS485 Modbus RTU communication
    Inputs1、IN1 input (start analysis /flow control switch / water meter)
    2、IN2 input (reset device)
    Analysis PeriodTime interval measurement (10~360min)/external signal/flow signal
    Flushing Timesettable (15~1800s)
    Water Quality Requirementscolourless, no suspended matter, no air bubbles
    Temperatureambient: 5°C – 45°C, water sample: 5°C – 40°C
    Humidity20 – 90% RF, indoor installation
    Pressureca. 0.5 – 5 bar (max.) (1 – 2 bar recommended)
    Water Inlet/Outlet Connection6 mm OD hose
    Dimensions/Weight300x300x200mm (WxHxD), ca. 4Kg
    MountingWall mounted
  • Order No.

    Description

    33-3400-00

    PACON 3400 Online Total Iron Analyzer

    50-3400-01

    FE3401 Low range Iron ion reagents

    50-3400-02

    FE3402 High range Iron ion reagents

     

    50-5000-10

    Spare Parts Kit Including pump head

    (including pump tube), all seals, stirrups, 

    connection pipe of reagent bottle,

    recommend to exchange every year.


Datasheet Download






Online Message
We will manage and protect the privacy of the personal information
  • Name
  • E-Mail
  • Telephone
  • Message Content

footer_background
Service Repair Form
Tell us what issue you encountered, including any steps or symptoms.
  • Name
  • Email
  • Phone
  • Fault Description

If you are interested in our products, you can choose to leave your information here, and we will be in touch with you shortly.